No study has so far compared Amulet with the new Watchman FLX in terms of residual left atrial appendage (LAA) patency or clinical outcomes in patients undergoing percutaneous LAA closure (LAAC). In the investigator-initiated SWISS APERO trial, patients undergoing LAAC were randomized (1:1) open-label to receive Amulet or Watchman 2.5 or FLX (Watchman) across 8 European centres. The primary endpoint was the composite of justified crossover to a non-randomized device during LAAC procedure or residual LAA patency detected by cardiac computed tomography angiography (CCTA) at 45 days. The secondary endpoints included procedural complications, device related thrombus (DRT), peridevice leak (PDL) at transesophageal echocardiography (TEE) and clinical outcomes at 45 days...