As low as diagnostically acceptable dose imaging in maxillofacial trauma: a reference quality approach

As low as diagnostically acceptable dose imaging in maxillofacial trauma: a reference quality approach

As-low-as-diagnostically-acceptable (ALADA) doses are substantially lower than current diagnostic reference levels. In this study, to improve dose management, a reference quality approach was tested in which phantom quality metrics of a clinical ALADA dose reference protocol were used to benchmark potential ALADA dose protocols for various scanner models. 36 different image series were analysed in the phantom study. Based on the phantom quality metrics, potential ALADA dose protocols at 1.73–1.92 mGy were selected. Compared with the reference images, the selected protocols showed non-inferiority in the detection and classification of maxillofacial fractures and non-inferior CNR of orbital soft tissues in the cadaver study. Based on the study findings, it was concluded that reference quality metrics from clinical ALADA dose protocols might be used to guide the selection of potential ALADA dose protocols for different CT scanners.