The aim of this study was to validate two new whole‐room indirfect calorimeters according to Room Indirect Calorimetry Operating and Reporting Standards (RICORS 1.0). Technical validation showed high reliability and recovery rates for VO2 (0.75% and 100.8%, respectively), VCO2 (0.49% and 100.6%), and EE (0.54% and 98.2%). Biological validation revealed CV and MDC for active conditions of 1.4% and 4.3% for 24‐hour EE, 1.7% and 5.9% for SMR, and 30.2% and 38.4% for TEF, as well as 5.8% and 10.5% for PAEE, respectively. The precision of 24‐hour EE and SMR was high, whereas it was lower for PAEE and poor for TEF.