Side‐by‐side catheterization with two catheters for intravesical and abdominal pressure during full cystometry in 36 prospectively recruited patients with analysis of mean and absolute differences at urodynamic events and post hoc in‐depth signal analysis comparing the full pressure traces of both systems. The mean pressure differences at urodynamic events between air‐filled and water‐filled systems are small, however, with a large variation, without a systematic difference. The majority of the intersystem differences are significantly larger than 5 cmH2O. Further analysis showed that urodynamic event pressure differences of both systems at the start of the test were carried forward throughout the remainder of the test without subsequent or additional tendency to differ...