This study was designed to investigate the influence of diameter reductions on the stress distribution around root‐analogue implants via 3D finite element analysis. Four root‐analogue implant models with different diameter reductions (0, 1, 2, or 3 mm), a traditional threaded implant and congruent bone models were created through reverse engineering. A 100‐N force was applied parallel with and in a 45° angle to the implant axis, respectively. The stress concentration in the labial neck area around implants with 1–2 mm diameter reduction was lower than seen with no reduction. When the implant diameter was reduced by 3 mm, there were obvious stress concentrations in both implant and bone (the maximum stress was 206 and 111 MPa, respectively)...