Reverse triggering dyssynchrony is a common issue in sedated critically ill patients on mechanical ventilation. It occurs when a patient's intrinsic respiratory rhythm is entrained by periodic mechanical insufflations from the ventilator, causing their inspiratory efforts to start after passive insufflation. Diagnosing reverse triggering can be challenging but can be detected by inspecting ventilator waveforms or using more sophisticated methods such as the electrical activity of the diaphragm or esophageal pressure. The occurrence of reverse triggering may have clinical consequences depending on the magnitude of the inspiratory effort and can cause breath-stacking and loss of protective lung ventilation. Little is known about managing patients with reverse triggering, but available evidence can guide management based on physiological principles.