08Oct 2022
Typical value ranges and signal patterns in patients with neurogenic bladder: Quality control in urodynamics using an air‐charged catheter system

Typical value ranges and signal patterns in patients with neurogenic bladder: Quality control in urodynamics using an air‐charged catheter system

To establish typical value ranges (TVRs) of the air‐charged catheter (ACC) system and analyze the distinct signal patterns (TSPs) of cough under different bladder volumes for quality control of a urodynamic study using the ACC system. The 50% range of the initial resting pves, pabd, and pdet in the supine and sitting positions were 7–15, 7–14, and 0–0 cmH2O, and 24–33, 24–33, and 0–0 cmH2O, respectively. TVRs of the initial resting state in patients with NB was established to guide the ACC system's quantitative quality control. In addition, the TSPs of the pdet cough signal under different bladder volumes and postures were described, which could be used for qualitative quality control of the ACC system.